45
Troubleshooting
SIGMA Manual Number 41019
Revision AA
402-411
NVM write error: x (1-
10)
NVM write failed
412
User NVM CRC Error
NVM CRC check failed (user data)
413
Cal NVM CRC Error
NVM CRC check failed (cal data)
414
Wrong NVM structure
NVM version # doesn’t match
expected version#
500
Red. Calc: Unknown
501
Red. Calc: Rate
Rate calculation data error
502
Red. Calc: Bolus Rate
Bolus rate calculation data
503
Red. Calc: Set Mech
Set mechanism state calculation data
504
Red. Calc: Baseline poly
Baseline polynomial data
505
Red. Calc: Temp poly
Temperature polynomial data
506
Red. Calc: Startup
Startup ln function data
507
Red. Calc: Downstream
Downstream algorithm
508
Red. Calc: Upstream
Upstream data
510
Cannot calc drug lmts
Unable to calculate drug limits
511-529
Improper Units: 1-19
Unimplemented rate/conc units
combination
530
Flow Cal out of Range
Flow calibration out of range
540
Bad Tube Type
Tube type incorrect or not selected
601
Sharp Code Corrupted
Code CRC failed
602
CPLD Invalid
CPLD ID wrong
603
Timebase Invalid
Sharp or PIC osc. frequency is wrong
604
Sharp ALU Failure
ALU failure
Error
Code #
Displayed and
History Log
Language
Description