17
Configuration Options
SIGMA Manual Number 41019
Revision AA
Sensor Configuration
■
Downstream Occlusion Auto-Restart on/off - Turns on or off the Downstream Occlusion
Auto-Restart feature.
■
US /Air alarm confirmation on/off - When this selection is set to on, the user must acknowl-
edge the air-in-line alarm with a key-press before the RUN key will be accepted to resume
the infusion.
Calibration Menu
The Calibration Menu provides access to sensor calibration data. There are 4 sensors in this menu
and, the biomed personnel can access the sensor data in these menus.
NOTE:
Calibration of the device shall be performed by technicians that have been
formally trained by SIGMA personnel on Spectrum calibration procedures
ONLY.
Sensor Calibration
■
Upstream Sensor
■
View Calibration Values - Provides a utility to view the Upstream Sensor calibration
data that follows.
■
Upper Freq (kHz)
■
Upper Power Level
■
Upper Reading
■
Lower Freq (kHz)
■
Lower Power Level
■
Lower Reading
■
Downstream Sensor
■
View Calibration Data - Provides a utility to view the Downstream Sensor calibration
data.
■
Current Reading
■
No-Tube Reading
■
Digit Pot Setting
■
Scale Factor
■
Scale Offset
■
Slide Clamp Sensor
■
View Calibration Data - Provides a utility to view the Slide Clamp Sensor and No Slide
Clamp calibration data.
■
Red %
■
Green %
■
Blue %
■
View Current Readings - Provides a utility to view the
current
Slide Clamp sensor data.
■
Red %
■
Green %
■
Blue %