41
Troubleshooting
SIGMA Manual Number 41019
Revision AA
Battery charger error codes may be cleared by performing the following steps:
1.
Remove the external power supply.
2.
Remove the battery for 1 minute.
3.
Install the battery again and connect the external power supply.
If the charge error repeats, return the pump to SIGMA for service.
System Error Code Descriptions
A series of error codes are generated by the software of the SIGMA Spectrum Infusion Pump, then
displayed on the pump. The following error codes are intended to help the biomedical engineer to
identify problems encountered during operation of the pump for assistance in repair of the product.
Error
Code #
Displayed and
History Log
Language
Description
100
PIC Error
Error of unknown type
101
PIC Msg CRC Error
CRC failure
102
PIC Not Responding
Not responding
103
PIC Msg Rcv CRC Er
Message receive CRC error
104
PIC Receive Ovrflow
Receive overflow
105
PIC Motor Error
Motor failure flag
106
PIC Index Error
Index failure flag
107
PIC CAM Error
CAM failure flag
108
PIC COM Error During
Run
COM failure flag
109
PIC Encoder Error
Encoder failure flag
110
PIC counts Per Cam Error
Counts per CAM failure flag
111
Pulse dmnd too high
Pulse demand too high for interval
mode
112
Pulse dmnd too low
Pulse demand too low for PWM mode
113
PIC ROM CRC Error
ROM CRC error
114
Sensor disparity (slow)
Motor is slowing down