10 Technical data
70 / 91
Version 06/2012, Rev. 1.11 of 06/04/2016 • sb
Translation of the original operating manual, part no. 0702202
Pos : 208 /100 Sig ma/110 BA Z entrif ugen Sig ma (Proj ekt e)/ 6-16KS_6- 16KHS/100 T ec hnisc he D at en100 T ec hnische D at en 6- 16KS============================== @ 29\ mod_1405327277386_68.doc x @ 199507 @ 1 @ 1
10 Technical data
Manufacturer
Sigma Laborzentrifugen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Type:
6-16KS
Connection requirements
Electr. connection:
Protection class:
IP code:
see name plate
I
20
Connected load (kVA):
Power consumption (kW):
Max. current consumption (A):
Input fuse (AT):
2.8
2.3
14.0 (at 220-240 V / 50 Hz)
16.0 (at 220-240 V / 50 Hz), temperature fuses
Performance data
Max. speed (rpm):
Max. capacity (ml):
Max. gravitational field (x g):
Max. kin. energy (Nm):
15,000
3,200
25,155
152,497
Other parameters
Time range:
Temperature range:
Storage locations:
10 sec – 99 h 59 min, short run, continuous run
-20°C to +40°C
60
Physical data
Height (mm):
Height with open lid (mm):
Width (mm):
Depth (mm):
Weight (kg):
485
990
790
711
158
Noise level (dB(A)):
< 68 (at maximum speed)
Special equipment:
Water cooling system
Tap connections:
Inlet pressure (bar):
Max. flow rate (m
3
/h):
Max. temperature at water inlet:
2 x ¾
1,5 bis 5,0
< 0,5
20°C