![Sigma 6-16KS Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/sigma/6-16ks/6-16ks_operating-manual_1267910051.webp)
Refrigerated Centrifuge Sigma 6-16KS
6 Using the centrifuge
Version 06/2012, Rev. 1.11 of 06/04/2016 • sb
51 / 91
Translation of the original operating manual, part no. 0702202
Restrictions
• Acceleration and deceleration curves can include intervals with a
positive slope as well as intervals with a negative slope and also
intervals with 0 slope.
• The slope of the curve intervals can be 1 min
-1
/sec (rpm/sec) minimum
and 1000 min
-1
/sec (rpm/sec) maximum.
• Quadratic curve intervals are only possible between 0 and 1000 min
-1
(rpm) maximum. If a final speed > 1000 min
-1
(rpm) is selected, this
interval will automatically become linear above 1000 min
-1
(rpm).
• The possible runtime results from the maximum possible speed
(depending on the rotor) and from the slope limitation.
Example 1: Start speed 0 min
-1
(rpm), final speed 100 min
-1
(rpm),
runtime 1 hour not possible, since the necessary slope < 0.03 min
-1
(rpm), which means that it is beyond the defined range.
Example 2: Start speed 0 min
-1
(rpm), final speed 15000 min
-1
(rpm),
runtime 10 sec. not possible, since the necessary slope is 1500 min
-1
(rpm), which means that it is beyond the defined range.
Pos : 143 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 144 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ Spinc ontrol S (Z ent7) 060-0030- 0020-0090 M enü Hilf e @ 27\ mod_1405320230870_68.doc x @ 193417 @ 4 @ 1
→
6.3.2.10 Help menu
The help function is symbolised by the " " symbol on the menu bar. It
provides a short description of the control elements of the selected option.
Activating and deactivating the help function
• Select the question mark on the menu bar and press the function knob.
• Quit the help function by selecting the question mark and by pressing
the function knob again.
Parameters can still be changed when the help function is activated.
Fig. 25: Menu "Help"
Pos : 145 /010 Uni vers almodul e/ Seit enwechs el @ 0\ mod_1202116244312_0.doc x @ 105 @ @ 1