Refrigerated Centrifuge Sigma 6-16KS
6 Using the centrifuge
Version 06/2012, Rev. 1.11 of 06/04/2016 • sb
33 / 91
Translation of the original operating manual, part no. 0702202
T
e
m
p
e
ra
tu
re
1
)
Sp
e
e
d
R
C
F
T
im
e
t
A
c
c
e
le
ra
ti
o
n
c
u
rv
e
2
)
D
e
c
e
le
ra
ti
o
n
c
u
rv
e
Pr
o
g
ra
m
°C
rpm
x g
min
no.
no.
22
3,750 4,544
7
20 (2 min)
12
1
Full blood
22
3,570 4,544
10
20 (2 min)
12
2
Full blood in 4fold bags
20
4,000 5,170
12
24 (3 min)
12
3
Plasma with buffy coat
22
3,750 4,544
9
20 (2 min)
12
4
Full blood
20
4,000 5,170
14
24 (3 min)
12
5
Plasma with buffy coat
22
1,094
384
7
21 (2 min)
11 or 2 min
6
Platelets with buffy coat
22
1,094
384
7
21 (2 min)
11 or 3.5min 7
Platelets with buffy coat
1)
only with special equipment heater
2)
Curves no. 20, 21 and 24 are freely programmable curves
NOTE
If the centrifuge does not reach the mentioned RCF values, the maximum
gravitational field must be set and the run time must be extended.
Pos : 119 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 120 /100 Sig ma/110 BA Z entrif ugen Sig ma (Proj ekt e)/ 6-16KS_6- 16KHS/060 Betri eb060- 0020-0020-0070 Zentrif ugati on mit Erdölgäs ern - nur KS @ 29\ mod_1405327245261_68. doc x @ 199465 @ 4 @ 1
6.2.2.7
Centrifugation with tubes for mineral oil
In the previous ASTM standard, a maximum gravitational field of 800 x g is
required for the centrifugation with tubes for mineral oil. Practically, it is
possible to reach higher RCF-values with the rotor 11165 and the buckets
13113, 13114 or 13116.
CAUTION
The required gravitational field must not be exceeded.
Valures above 800 x g increase the risk of glass breakage!
• Use sealing caps (part no. 17107).
Pos : 121 /010 Uni vers almodul e/ Seit enwechs el @ 0\ mod_1202116244312_0.doc x @ 105 @ @ 1