![Sigma 3-18KS Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/sigma/3-18ks/3-18ks_operating-manual_1267900039.webp)
Refrigerated Centrifuge Sigma 3-18KS
6 Using the centrifuge
Version 08/2012, Rev. 1.15 of 15/09/2017
• sb
39 / 89
Translation of the original operating manual, part no. 0701102
Pos : 134 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ Spinc ontrol S (Z ent7)/ 060- 0030-0020-0055 M enü Prozes sbi bliot hek @ 27\ mod_1405320209123_68.doc x @ 193276 @ 4 @ 1
6.3.2.6
Process library menu
The process library can be opened via the "
" symbol on the menu bar. It
provides the user with an overview of all of the stored programs as
processes.
The processes are listed with their name, RCF, runtime, and temperature
(exception: the "RAPID_TEMP" program will not be listed).
The order of the processes corresponds to their storage locations on the
program list. Empty storage locations will not be displayed. If more than 11
programs have been stored, the user can scroll through the list.
Fig. 15: Process library menu
Loading a process
• In the process library, select the desired process by turning the function
knob. Press the function knob in order to confirm the selection.
The process will be loaded and the "Standard" menu will be displayed.
Starting a process
• In the process library, select the desired process by turning the function
knob. Then, press the start button.
The process will be loaded and started. The "Standard" menu will be
displayed.
Pos : 135 /010 Uni vers almodul e/ Seit enwechs el @ 0\ mod_1202116244312_0.doc x @ 105 @ @ 1