![Sigma 3-18KS Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/sigma/3-18ks/3-18ks_operating-manual_1267900029.webp)
Refrigerated Centrifuge Sigma 3-18KS
6 Using the centrifuge
Version 08/2012, Rev. 1.15 of 15/09/2017
• sb
29 / 89
Translation of the original operating manual, part no. 0701102
Pos : 114 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0030 Eins et zen von Z ubehör 4plätzig bis 2-16P bis 3- 18KS @ 27\ mod_1405319205710_68.doc x @ 192507 @ 4 @ 1
6.2.2.3
Installation of accessories
• Only use inserts that are suitable for the rotor (see chapter 11.1 -
• All buckets of the swing-out roto need to be installed when spinning.
• Always load the axial symmetrical inserts/buckets of the rotors with the
same accessories and fill to avoid imbalance.
Centrifugation with different tube sizes
Working with different tube sizes is possible. In this case, however, it is
very important that axial symmetrical inserts are identical.
correct
incorrect
Fig. 6: Permissible and impermissible loading of a swing-out rotor with different tube sizes
(example illustrations)
Centrifugation with low capacity
• Install the tubes axial symmetrically so that the buckets and their inserts
are loaded evenly.
• It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 7: Permissible and impermissible loading of an angle rotor and a swing-out rotor
(example illustrations)