3.7
Safety devices
Pos : 61 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0010 D ec kel verri egelung @ 27\ mod_1405319036321_68.doc x @ 192171 @ 3 @ 1
3.7.1
Lid lock device
The centrifuge can only be started when the lid is properly closed. The
electrical lock must be locked. The lid can only be opened when the rotor
has stopped. If the lid is opened by way of the emergency release system
during operation, the centrifuge will immediately switch off and decelerate
brakeless. If the lid is open, the drive is completely separated from the
mains power supply, i.e. the centrifuge cannot be started (see chapter 7.1.1
- "Emergency lid release").
Pos : 62 / 010 U ni vers al module/ Leerz eile 8pt @ 12\mod_1337063062114_0. doc x @ 65687 @ @ 1
Pos : 63 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0020 Stillst ands über wachung @ 27\ mod_1405319037803_68. doc x @ 192185 @ 3 @ 1
3.7.2
Standstill monitoring system
Opening of the centrifuge lid is only possible if the rotor is at a standstill.
This standstill is checked by the microprocessor.
Pos : 64 / 010 U ni vers al module/ Leerz eile 8pt @ 12\mod_1337063062114_0. doc x @ 65687 @ @ 1
Pos : 65 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0030 Syst emkontroll e Spincontr ol S @ 27\ mod_1405319039294_68. doc x @ 192199 @ 3 @ 1
3.7.3
System check
An internal system check monitors the data transfer and sensor signals with
regard to plausibility. Errors are detected with extreme sensitivity and
displayed as error messages in a dialog box (see chapter 7.2 - "Table of
error codes").
Pos : 66 / 010 U ni vers al module/ Leerz eile 8pt @ 12\mod_1337063062114_0. doc x @ 65687 @ @ 1
Pos : 67 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0040 Sc hutzleit erpr üf ung Spinc ontrol L+S+ Eas y+Uni vers al @ 27\ mod_1405319043129_68. doc x @ 192241 @ 3 @ 1
3.7.4
Earth conductor check
For the earth conductor check, there is an equipotential bonding screw on
the rear panel of the centrifuge (see chapter 2.1.1 - "Functional and
operating elements"). An earth conductor check can be carried out by
authorized and specialized personnel using a suitable measuring
instrument. Please contact the service department (see chapter 7.3 -
"Service contact").
Pos : 68 / 010 U ni vers al module/ Leerz eile 8pt @ 12 \mod_1337063062114_0. doc x @ 65687 @ @ 1
Pos : 69 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0050 U nwuchtüberwac hung Spi ncontr ol S @ 27\ mod_1405319045827_68.doc x @ 192269 @ 3 @ 1
3.7.5
Imbalance monitoring system
A dialog box may pop up or emit a sound signal in order to indicate that the
centrifuge is in the inadmissible imbalance range. If the rotor is loaded
unevenly, the drive will be switched off in the acceleration phase or during
the run.
Pos : 70 / 010 U ni vers al module/ Leerz eile 8pt @ 12\mod_1337063062114_0. doc x @ 65687 @ @ 1
Pos : 71 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0060 T emperatur über wac hung @ 27\ mod_1405319048438_68. doc x @ 192297 @ 3 @ 1
3.7.6
Temperature monitoring system
If the temperature inside the rotor chamber rises above 50°C, the drive
system will be switched off automatically. The centrifuge cannot be
restarted until it has cooled.
Pos : 72 / 010 U ni vers al module/ Leerz eile 8pt @ 12 \mod_1337063062114_0. doc x @ 65687 @ @ 1
Pos : 73 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0070 R ot orüberwac hung @ 27\ mod_1405319049932_68.doc x @ 192311 @ 3 @ 1
3.7.7
Rotor monitoring system
When a rotor number and, if applicable, a bucket number are selected, the
computer will automatically check whether the entered speed or the
entered gravitational field are permissible for the selected rotor.