6 Using the centrifuge
28 / 89
Version 08/2012, Rev. 1.15 of 15/09/2017
• sb
Translation of the original operating manual, part no. 0701102
CAUTION
When using rotors for microtiter plate formats:
Ensure that the plate holders are inserted together with the plates into the
buckets.
NOTE
The lid screw serves for the fastening of the lid onto the rotor only, not for
the fastening of the rotor onto the motor shaft.
• Follow the safety instructions and hazard warnings (see chapter 3 -
Pos : 111 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 112 /100 Sig ma/110 BA Z entrif ugen Sig ma (Proj ekt e)/ 3-18KS-3- 18KH S/ 060 Betrieb/060-0020-0020- 0020 3-18KS_KH S Ei ns etz en von Wi nkelrot oren mit her metisch verschli eßbarem Dec kel @ 29\ mod_1405321818963_68. doc x @ 197771 @ 4 @ 1
6.2.2.2
Installation of angle rotors with a hermetically sealed lid
• Screw the rotor cover onto the rotor and tighten it.
• Lower the rotor with the cover onto the motor shaft .
• Insert the rotor tie-down screw into the motor shaft. Tighten the rotor tie-
down screw with 7.5 Nm using the supplied rotor wrench so that the
spring washer assembly is compressed tightly.
• The rotor can also be used without a cover.
• The rotors can be installed or removed with a closed lid after loosening
the rotor tie-down screw.
• Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
1
Rotor tie-down screw
2
Lid
3
Rotor
Fig. 5: Angle rotor with a hermetically sealed lid