Maintenance and Service
50
Version 03/2010, Rev. 1.7 of 27/01/2012
Translation of the original operating manual
Accessories
max. temp.
(°C)
min. time max. time max.
cycles
Glass tubes
134-138
3
40
-
Polycarbonate tubes
115-118
30
40
20
Polypropylene copolymer tubes
115-121
30
40
30
Teflon tubes
134-138
3
5
100
Aluminium rotors
134-138
3
20
-
Polycarbonate/Polyallomer lids for angle rotors
115-118
30
40
20
Polysulfone lids for angle rotors
134-138
3
5
100
Aluminium buckets
134-138
3
20
-
Polysulfone caps for buckets
134-138
3
5
100
Rubber adapter
115-118
30
40
-
Round carriers made of polypropylene
115-118
30
40
-
ditto, made of polyallomer and polycarbonate
115-118
30
40
-
Rectangular carriers made of polypropylene
115-118
30
40
-
ditto, made of polyallomer and polycarbonate
115-118
30
40
-
Fig. 8.1: Autoclaving table
8.3 Service
In the event of service work that requires the removal of the panels,
there is a risk of electric shock or mechanical injury. Only qualified
specialist personnel is authorised to perform this service work.
The centrifuge is subject to high mechanical stress. In order to be
able to withstand this high level of stress, high-quality components
were used during the production of the centrifuge. Nevertheless, wear
cannot be excluded and it may not be visible from the outside.
Especially the rubber parts that are – among other things – part of the
motor suspension, are subject to ageing.
This is why we recommend having the centrifuge checked by the
manufacturer during an inspection once per year in the operating
state and once every three years in the dismantled state. Rubber
parts should be replaced after three years or a maximum of 15,000
cycles.