![Sigma 134226 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/sigma/134226/134226_operating-manual_1267882034.webp)
2-16KC
Using the Centrifuge
34
Version 03/2010, Rev. 1.7 of 27/01/2012
Translation of the original operating manual
An entry will interrupt the rapid cooling program. There will be no
message when the set temperature is reached. If no new runtime is
entered, the centrifuge will continue to run infinitely (continuous run).
Start Delay
If the start delay is activated, the centrifuge will not start until the
preset time has elapsed.
Automatic Lid Opening Function
The Automatic-Lid-Open function must be activated so that the lid
opens automatically when the rotor is at a standstill.
When the lid is open, the cooling is not active. Samples may warm
up!
6.3.3.7 Configuration Menu
In the configuration mode, the default settings can be customised and
the data can be read off.
Fig. 6.16: Configuration menu
Code
In order to prevent any unauthorised use of the centrifuge, the
following functions can be blocked:
Saving of programs
Changing of parameters
Loading of programs
Start key
Blocking a function:
Select the function that is to be blocked.
Select the button "Activate code".
Enter a four-digit code and confirm the entry.
For safety reasons, the code must be entered a second time.
The blocking is now active.
If changes are made to a blocked function, the system will ask for the
code prior to executing the change.