Malfunctions ans Error Correction
42
Version 03/2010, Rev. 1.7 of 27/01/2012
Translation of the original operating manual
7
Malfunctions and Error Correction
7.1 General Malfunctions
Malfunctions are indicated by a dialogue box. If the sound signal is
activated, it sounds when the error message is displayed.
Eliminate the source of the problem (see tables in fig. 7.1 and 7.3).
Acknowledge the error messages by pressing the knob.
Error messages that are caused by the power circuit board are
marked with a “P” in front of the error code.
Type of error
Possible reason
Correction
No power in the mains supply
Check fuse in the mains supply
Power cord is not plugged in
Plug in power cord correctly
Fuses have tripped
Reactivate temerature fuse (see
5.4.1 "Connection")
Mains power switch off
Switch mains power switch on
No indication on the display
Lid is not closed correctly
Close lid.
Centrifuge cannot be started:
Start key LED is not illuminated
Several
Power off/on. If the error occurs
again, contact service
Centrifuge cannot be started:
Lid key LED flashes
The lid lock is not closed
correctly
Open and close lid. If the error
occurs again, contact service
Brief mains power failure
Press start key in order to
restart the centrifuge
Centrifuge decelerates during
operation
System error
Power off/on. If the error occurs
again, contact service
- improper loading
- Centrifuge is inclined
- Drive problem
- centrifuge was moved
during run
Balance loading and restart the
centrifuge. If the error occurs
again, contact service
Centrifuge decelerates during
operation, imbalance dialogue
box is displayed
- ungreased load-bearing bolts
Clean and grease load-bearing
bolts
Lid lock has not released
Unlock the lid manually (see
7.1.1) and contact service
Lid cannot be opened
Lid seal sticks
Clean the lid seal and apply
talcum powder
Temperature value cannot be
reached
Condenser dirty
Clean the condenser. If the error
occurs again, contact service
Fig. 7.1: Possible error causes