6.2.4
Lid cannot be opened
−
When first trying to open the lid,
the locks are not released. Lid
key LED flashes.
•
Close the lid again. Press down both
sides of the lid until the locks audibly
lock. Open the lid again. If the error
occurs again, unlock the lid manually
(see 6.2.5 "Emergency lid release",
page 42) and call the service.
−
The lid cannot be opened,
although both locks have audibly
unlocked.
•
Check/clean the lid seal. Apply talcum
powder to the seal to avoid sticking.
6.2.5
Emergency Lid Release
In the event of a power supply failure, it is possible to manually open the
centrifuge lid using a stopper that is secured with a screw. It is located on the
front of the bottom panel.
•
Switch off the mains power switch and disconnect the power cord from the
socket.
•
Unscrew the screw in the stopper.
•
Remove the stopper, e.g. with a screwdriver.
•
Pull the visible string in order to release the lid. Then, reinsert the stopper into
the opening.
•
Reinsert the screw centrally.
The lid may only be unlocked and opened when the rotor is at a standstill.
If the lid is opened by way of the emergency lid release during operation, the
centrifuge will be switched off immediately and decelerate brakeless.
Screw
Stopper
String
Fig. 6.1: Screw for the emergency lid release
Operating Manual SIGMA 1-6P
page 42 of 67
05/2007
Содержание 1-6P
Страница 1: ...Laboratory Centrifuge 1 6P Operating Manual From Serial No 119956 Please retain for later use...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 14: ......
Страница 54: ...Operating Manual SIGMA 1 6P page 54 of 67 05 2007 Abb 11 1 Drehzahl Schwerefeld Diagramm...
Страница 56: ......
Страница 58: ......
Страница 60: ......
Страница 64: ......