5.3.5 Program
The program is used to save or load certain recurrent settings of the centrifuge.
This saves time and prevents typing errors. Ten different programs (0-9) can be
saved and called up. This is only possible at a standstill.
5.3.5.1 Saving the Current Settings
To save the current settings:
•
Turn the left rotary knob until the “save” cursor appears on the display (Fig.
5.14).
•
Select the option by pressing the right rotary knob. The “save” cursor will now
flash.
•
Turn the right rotary knob until the desired program number is displayed.
Empty storage location numbers flash.
•
Press the right rotary knob to confirm the input.
The settings are now saved under this program number (program number 8 in
the example).
Fig. 5.14: Saving programs;
here program number 8
5.3.5.2 Calling Up Stored Programs
If necessary, the saved programs can be called up:
•
Turn the left rotary knob until the “run” cursor appears on the display (Fig
5.15).
•
Select the option by pressing the right rotary cursor. The “run” cursor will now
flash.
•
Turn the right rotary knob until the desired program number is displayed.
•
Press the right rotary knob to confirm the input.
The saved program settings will be adopted.
Fig. 5.15: Calling up stored programs;
here program number 8
Operating Manual SIGMA 1-6P
page 38 of 67
05/2007
Содержание 1-6P
Страница 1: ...Laboratory Centrifuge 1 6P Operating Manual From Serial No 119956 Please retain for later use...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 14: ......
Страница 54: ...Operating Manual SIGMA 1 6P page 54 of 67 05 2007 Abb 11 1 Drehzahl Schwerefeld Diagramm...
Страница 56: ......
Страница 58: ......
Страница 60: ......
Страница 64: ......