![Sigma 1-15PK Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/sigma/1-15pk/1-15pk_operating-manual_1267874048.webp)
Version 08/2008, Rev. 1.7 - sb
page 48 of 75
English translation
6.2.8 Error
Codes
Error no.
Kind of error
Measure
Note
1-9
System error
allow to slow down
power off/on
All these errors stop
the centrifuge or cause
it to slow down.
10-19
Speedometer error
allow to slow down
power off/on
20-29
Motor error
power off/on
ensure ventilation
30-39
EEPROM error
allow to slow down
power off/on
With errors 34,35,36,
the centrifuge will stop.
With errors 37,38,
error message only.
40-45 Temperature
error
(only for refrigerated
centrifuges)
allow to slow down
power off/on
allow to cool down
provide better ventilation
46-49
Imbalance error
allow to slow down
power off
eliminate the imbalance
50-59
Lid error
press lid key
close lid
remove foreign matter from
the opening of the lid lock
device
With error 50 and 51,
the centrifuge will stop.
60-69
Process error
allow to slow down
power off/on
With error 60,
message “power
failure during run”.
With error 61,
message “stop after
power on”.
70-79
Communication error
allow to slow down
power off/on
80-99
Parameter error
allow to slow down
provide for better ventilation
power off/on
With error 83, error
message only.
)
Attention!
If it is impossible to eliminate the errors, please call the service!
Fig.6.2: Error Codes
Содержание 1-15PK
Страница 1: ...Operating Manual Refrigerated Centrifuge 1 15PK from serial no 120506 Please retain for later use...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 12: ......
Страница 14: ......
Страница 61: ...Version 08 2008 Rev 1 7 sb page 61 of 75 English translation Fig 11 1 Speed gravitational field diagram...
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 72: ......