7 Malfunctions and error correction
Version 05/2010, Rev. 2.12 of 15/08/2016 • sb
39 / 70
Translation of the original operating manual, part no. 0700102
Pos: 152 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 070 St öru nge n un d Fehle rsuc he/0 70 St öru nge n un d F ehle rsuc he= == === == == == === == == === == == === @ 2 8\m od_ 1405 320 576 072 _68. docx @ 19 466 3 @ 1 @ 1
7 Malfunctions and error correction
Pos: 153 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 070 St öru nge n un d Fehle rsuc he/0 70- 001 0 Allge mein e Stö rung en - Basic (au ßer 2- 7) @ 28\ mod _14 053 205 774 37_ 68.d ocx @ 194 677 @ 2 @ 1
7.1 General malfunctions
Malfunctions are indicated by error messages with a number in the
speed/RCF display.
In the event of a fatal error (e.g. a defective lid lock), a certain safety time
will be counted down on the display. During this time, “ERR” and “SAFE”
flash alternately on the display. When the time is up, "OFF" will be
displayed.
WARNING
Do not switch the centrifuge off until “OFF” is displayed! This is necessary
in order to ensure that the rotor is at a complete standstill.
• Eliminate the source of the problem (see tables below).
• Acknowledge the error messages with the lid key.
Type of error
Possible reason
Correction
No indication on the display
No power in the mains supply
Check fuse in the mains supply
Power cord is not plugged in
Plug in power cord correctly
Mains power switch off
Switch mains power switch on
Lid is not closed correctly
Close the lid
Centrifuge cannot be started: The set
speed value is displayed in an
unchanged manner
Several
Netz aus/ein. Falls sich der Fehler
wiederholt, Service verständigen
The lid lock is not closed correctly
Open and close lid. If the error
occurs again, contact service
Centrifuge decelerates during
operation and displays an error from 1
to 18 after powering on
Several
Power off/on. If the error occurs
again, contact service
Centrifuge decelerates during
operation and displays error 19 after
powering on
Several
Quit by pressing the lid key
Temperature value cannot be reached
(only for refrigerated centrifuges)
Condenser dirty
Contact service