Refrigerated Centrifuge Sigma 1-14K
6 Using the centrifuge
Version 05/2010, Rev. 2.12 of 15/08/2016 • sb
29 / 70
Translation of the original operating manual, part no. 0700102
WARNING
Never hold the rotor on the outside when inserting and removing it.
Crushing hazard!
• Hand-tighten the rotor fastening nut clockwise (corresponds to 1 Nm).
• Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
Pos: 104 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 105 /10 0 Sigm a/11 0 BA Z entrif uge n Sigm a (Pr ojekt e)/1 -14 _1- 14K/0 61 Bet rieb/ 060 -00 20- 002 0-0 020 Einsetz en d es Mik roh äma tokrit telle rs 1- 14_ 1-1 4K @ 2 8\m od_ 140 532 0812 176 _68 .docx @ 1 9548 9 @ 4 @ 1
6.2.2.3 Installation of the microhaematocrit rotor
Fig. 9: Microhaematokrit rotor (part no. 11026) with reader (part no. 17028)
• Unscrew the rotor fastening nut from the motor shaft (anti-clockwise).
• Lower the microhaematocrit rotor with its central bore straight down
onto the motor shaft.
• Hand-tighten the rotor fastening nut clockwise (corresponds to 1 Nm).
The rotor fastening nut and conical spring washer must face the
microhematocrite rotor (i.e. the writing must be shown on top).
1
Conical spring washer
Fig. 10: Rotor fastening nut from above (A) and from below (B)
• Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!