Select Waveform Interpolation Method
Under real-time sampling, the oscilloscope acquires the discrete sample values of the
waveform being displayed. In general, a waveform of dots display type is very difficult to
observe. In order to increase the visibility of the signal, the digital oscilloscope usually
uses the interpolation method to display a waveform.
Interpolation method is a processing method to ―connect all the sampling points‖, and
using some points to calculate the whole appearance of the waveform. For real-time
sampling interpolation method is used, even if the oscilloscope in a single captures only a
small number of sampling points. The oscilloscope can use interpolation method for filling
out the gaps between points, to reconstruct an accurate waveform.
Press the
Acquire
button on the front panel to enter the ACQUIRE Function menu; then
press the
Sinx/X
softkey to select
Sinx
or
X
.
X:
In the adjacent sample points are directly connected on a straight line. This
method is only confined to rebuild on the edge of signals, such as square wave.
Sinx:
Connecting the sampling points with curves has stronger versatility. Sinx
interpolation method uses mathematical processing to calculation results in the actual
sample interval. This method bending signal waveform, and make it produce more
realistic regular shape than pure square wave and pulse. When the sampling rate is 3
to 5 times the bandwidth of the system. Recommended sinx interpolation method.
Figure 11 Display Type Set to Dots
www.biall.com.pl
Содержание SDS1102X
Страница 2: ...www biall com pl...
Страница 19: ...SIGLENT SDS1000X User Manual XVII www biall com pl...
Страница 20: ...www biall com pl...
Страница 61: ...SIGLENT SDS1000X User Manual 41 Figure 12 x Interpolation Figure 13 Sinx Interpolation www biall com pl...
Страница 89: ...SIGLENT SDS1000X User Manual 69 Figure 28 Relative Window Trigger www biall com pl...
Страница 91: ...SIGLENT SDS1000X User Manual 71 Figure 29 Interval Trigger www biall com pl...
Страница 95: ...SIGLENT SDS1000X User Manual 75 Figure 32 Runt Trigger www biall com pl...
Страница 109: ...SIGLENT SDS1000X User Manual 89 Figure 44 SPI Trigger www biall com pl...