SIGLENT
SDG1000X User Manual
21
To Set the Offset
The offset setting range is limited by the ―Load‖ and ―Amplitude/HighLevel‖
settings. For detailed information, please refer to ―SDG1000X Datasheet‖. The
default value is 0Vdc.
1. Press Waveforms
→ Sine → Offset, to set the offset.
The offset shown on the screen when the instrument is powered on is the
default value or the set value of last power down. If you want to set the
waveform by low level, press the Offset/LowLevel key again, to switch into the
low level parameter (the current operation is displayed in inverse color).
2. Input the desired offset.
Use the numeric keyboard to input the parameter value directly, and press the
corresponding key to select the parameter unit. Or use the arrow keys to
select the digit to edit, and then use the knob to change its value.
Figure 2-4 Setting the Offset
To Set the Phase
1. Press Waveforms
→ Sine → Phase, to set the phase.
Содержание SDG1000X Series
Страница 2: ......
Страница 44: ...SIGLENT 34 SDG1000X User Manual Figure 2 15 Setting the Mean...
Страница 63: ...SIGLENT SDG1000X User Manual 53 The methods of setting the parameters of DSB AM are similar to AM...
Страница 119: ...SIGLENT SDG1000X User Manual 109 Figure 2 65 Phase Deviation Interface...
Страница 134: ...SIGLENT 124 SDG1000X User Manual Figure 3 1 Generate a Sine Waveform...
Страница 136: ...SIGLENT 126 SDG1000X User Manual Figure 3 2 Generate a Square Waveform...
Страница 148: ...SIGLENT 138 SDG1000X User Manual Figure 3 9 Generate an AM Modulation Waveform...
Страница 150: ...SIGLENT 140 SDG1000X User Manual Figure 3 10 Generate a FM Modulation Waveform...
Страница 152: ...SIGLENT 142 SDG1000X User Manual Figure 3 11 Generate a PM Modulation Waveform...
Страница 154: ...SIGLENT 144 SDG1000X User Manual Figure 3 12 Generate a FSK Modulation Waveform...
Страница 156: ...SIGLENT 146 SDG1000X User Manual Figure 3 13 Figure 3 13 Generate an ASK Modulation Waveform...
Страница 158: ...SIGLENT 148 SDG1000X User Manual Figure 3 14 Generate a PSK Modulation Waveform...
Страница 160: ...SIGLENT 150 SDG1000X User Manual Figure 3 15 Generate a PWM Modulation Waveform...
Страница 162: ...SIGLENT 152 SDG1000X User Manual Figure 3 16 Generate a DSB AM Modulation Waveform...