4.4
Quick start V12
1.
Connect your V12 Geophone with your INFRA Datalogger
by using any of the available INFRA Sensor Cables.
2.
Start the Datalogger and wait for the initialization process to
complete.
3.
Under the menu
Nodes
you’ll find your V12 unit as three
separate sensors:
GEO V12V-
<serial.no>
,
GEO V12L-
<serial.no>
,
GEO V12T-
<serial.no>
,
where GEO is the Customer string and
<serial.no>
is the
serial number.
4.
For each of the three sensors, set Standard (S01, S02, etc.).
Then for the chosen standard select
• Interval time
• Record time
• Threshold (level and ON/OFF).
It is also possible to change the Customer string which in
some standards is used to select weighting type.
Note! When performing a test or reference measurement,
make sure that the threshold is set to OFF. Otherwise
unnecessary transients will be recorded.
5.
Activate REG ON, and the measurement will be initiated.
6.
Note!
First registration of interval data is delayed to
synchronize to the local time. The delay may last up to one
interval period. First presentation of data will then be delayed
up to two interval periods. Transient triggering and readout
of Live-values are available directly after activation of REG
ON.