3.16.4.
Example 4: Measure 4-pair PSD and SNR of Calibration Partner linked at 1000Base-T
This task is performed using the
Link Config
and
Meters
menu. Port 1,1 should be connected to Port 1,2 using the
Sifos furnished patch cable.
Hint: If test port is Slot 1, Port 1, then Calibration Partner port is Slot 1, Port 2.
Menu
Task
Screen Shot
Port Select
Select the port to configure.
Click on
Port 2
in
Slot 1
on the
Slot Port
Select
menu.
Link Config
Reset calibration partner port to default
TEST PHY configuration
Select the
Link Config
Tab
Click
Reset Test PHY
Port Select
Select the test port to use for
measurements
Click on
Port 1
in
Slot 1
on the
Slot Port
Select
menu.
Link Config
Configure the test port speed, polarity and
duplex settings.
Select the
Link Config
Tab.
Select
Nominal Fixed
under
Test PHY
Click
Apply Rx Level
Select
1000BaseT
under
Link Speed
Select
Automatic
under
Link Duplex
Select
Automatic
under
Link Polarity
Click
Arm Configuration
Click
RE-LINK
Meters
Measure SNR.
Select the
Meters
Tab
Select
1000
under
Link Speed
Select
SNR
under
Meter Type
Select
All
under
Pair(s)
Click
Measure
Meters
Measure PSD.
Select the
Meters
Tab
Select
1000
under
Link Speed
Select
PSD
under
Meter Type
Select
All
under
Pair(s)
Click
Measure
Expect
–2.6dB
&
Flat Frequency
Response
Содержание PhyView PVA-3000
Страница 22: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 22 ...
Страница 70: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 70 ...
Страница 108: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 108 ...
Страница 124: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 124 ...