Sierra Instruction Manual
InnovaSonic
®
203 Economical
15
T r a n s d u c e r T y p e [ 2 3
0 . St a n d a r d
Diag./< Key
Press this key to display System Error Code. The
function is the same with Window M08.
2.2.
Menu System Examples
Example, pipe has diameter of DN200, measuring water, Pipe Material is carbon steel, No Liner
Step 1. Pipe outside diameter:
Press
keys to enter menu M11,
enter the pipe outside diameter of
, then press the
key to
confirm.
Step 2. Pipe wall thickness
Press
key to enter menu M12,
enter the pipe wall thickness of
, then press
the
key to confirm. Note: If this value is a
decimal less than 1.00, it must be preceded by a
zero. Example: .375 must be entered as 0.375.
Step 3. Pipe material
Press
keys to enter menu M14,
press the
key to bring up the scroll down
options, press the
or
key to scroll
to the correct pipe material (in this case carbon
steel), then press the
key to confirm.
Step 4. Transducers type
Although other types are available, select
STANDARD transducers.
Step 5. Transducer mounting methods
Press
key to enter menu M24,
press the
key, press the
or
key to select transducer-mounting method, and
press the
key to confirm. IN MOST
CASES, start with the V method and switch to
the Z method if signal strength/quality is
insufficient.
Step 6. Transducer spacing
Press
key to enter menu M25,
THE METER WILL CALCULATE THE
CORRECT
SPACING
FOR
THE
MOUNTING METHOD selected in menu 24
above. Install as directed.
* R- - - - - - - - -
Sy s t e m No r ma l
Pi p e Ou t e r Di a me t e r
2 0 0 mm
Pi p e Wa ll T h i c k n e s s
6 mm
Pi p e Ma t e r i a l [ 14
0 . Ca r b o n St e e l
T r a n s d u c e r Mo u n t i n g
0 . V
T r a n s d u c e r Sp a c i n g
15 9 .8 6 mm