Sierra Instruments Instruction Manual
Smart-Trak
®
2 Series 100
46
From this screen you can change the zero flow bridge differential value to match
the zero flow conditions of your application. Mount (or place) the instrument where
it is intended to be used, minding orientation, tilt, etc.; The accuracy of this function
is dictated on how close to operating conditions the unit is when the meter is
zeroed. The display will show:
Zero Meter
Confirm zero flow
Then press enter
Once you push enter, the unit will monitor the sensor bridge differential value. The
display will show:
Zero Meter
Confirm zero flow
New Zero
Upon pressing the enter key again, the unit will record the bridge differential value
to the zero flow condition, leaving you with a freshly “zeroed” instrument! You can
make additional changes by using the left and right arrow keys to move to other
Lower Level Screens. You may also choose to press the escape key to return to the
Upper Level Screens.
Change Password Screen
If you are already in the Lower Level Screens, push the right or left arrow until you
reach the Change Password screen. To get to this screen
at any time: Press
Escape—Press Enter—enter password and Press Enter. Then, press the Right or
Left arrow until you reach this screen
.
From this screen you can change the instrument password from the factory default
to any four-digit password of your choice. The display will show the factory default
password of four zero’s: “0000”. The display will show:
Change Password
0000
To make a change, press the enter key. The first digit will begin to blink. Use the
up and down arrows to modify this digit or the left and right arrows to choose
another digit. When you are satisfied, press enter. The display will show:
Change Password
0000