
4BWEIGHING FUNCTIONS
Bit
Meaning
15 14 13 12 11 10 9
8
7
6
5
4
3
2
1
0
Char. value of the load cell
0
0
≤
1 mV/V
0
1
≤
2 mV/V
(default)
1
0
≤
4 mV/V
1
1
Reserve
Decimal
place
for
remote
display
0
0
0
Decimal place 0
(default)
0
0
1
Decimal place 1
0
1
0
Decimal place 2
0
1
1
Decimal place 3
1
0
0
Decimal place 4
1
0
1
Decimal place 5
Filter setting
0
0
0
No
filtering
0
0
1
Cutoff frequency: 5 Hz
0
1
0
Cutoff frequency: 2 Hz
(default)
0
1
1
Cutoff frequency: 1 Hz
1
0
0
Cutoff frequency: 0.5 Hz
1
0
1
Cutoff frequency: 0.2 Hz
1
1
0
Cutoff frequency: 0.1 Hz
1
1
1
Cutoff frequency: 0.05 Hz
Mean value filter
0
No mean value filter
(default)
1
Mean value filter switched on (30x20ms)
Memory goal
(only relevant for limit values and zero set value via
set to zero command)
0
Carry over into EEPROM
(default)
1
Keep in RAM only
Operational state - channel
0
Channel switched on
(default)
1
Channel switched off
(default channel2 of U1)
Reserve
0
1
Weight unit
(only relevant for SIWATOOL, is only saved in SIWAREX U)
0
0
0
‚ ‚
(default)
0
0
1
‚ g ‚
0
1
0
‚ kg ‚
0
1
1
‚ t ‚
1
0
0
‚ oz ‚
1
0
1
‚ lb ‚
1
1
0
, lbs‚
1
1
1
‚ kN ‚
Reserve
(Reserve, currently not used)
0
(default)
1
Figure 5-2
Parameter table for setting data
SIWAREX U
5-29