The Transducer Block
3.9 PR_CUST_LIN Block (PA Slot 4) Parameter List
SITRANS TH400 PROFIBUS® PA
40
Configuration Manual, 02/2007, A5E01039143-01
3.9.4
Custom Polynomial Linearisation, Parameter List
Parameter
Rel.
Index
PA
Description
Type
Store Size
byte
RO /
R/W Min. Max. Default
CUSTOM_TC_NAME
8
Name of Custom defined TC ( LIN_TYPE = 240)
OCTET_
STRING SRC 20
R/W
”Linear
TC; no
RJ”
CUSTUM_TC_POLY_COU
NT
9
Number of 4. order polynomial parts for Custom defined TC
Unsigned
8
SRC 1
R/W
5
CUSTOM_TC_MIN_IN
10
Minimum input limit in µV for Custom defined TC
Float
SRC 4
R/W
0
CUSTOM_TC_MIN_OUT 11
Minimum usable output value in °C of polynomial set for
Custom defined TC
Float
SRC 4
R/W
0
CUSTOM_TC_MAX_OUT 12
Maximum usable output value in °C of polynomial set for
Custom defined TC
Float
SRC 4
R/W
1500,00
CUSTOM_TC_POLY_1
13
Polynomial part 1 of Custom defined TC converting µV to °C.
Consisting of: maximum input value in µV, a4..a0 polynomial
coefficients.
6*Float
SRC 24
R/W
30000;
0; 0; 0;
0,01; 0
CUSTOM_TC_POLY_2
14
Polynomial part 2 of Custom defined TC converting µV to °C.
Consisting of: maximum input value in µV, a4..a0 polynomial
coefficients.
6*Float
SRC 24
R/W
60000;
0; 0; 0;
0,01; 0
CUSTOM_TC_POLY_3
15
Polynomial part 3 of Custom defined TC converting µV to °C.
Consisting of: maximum input value in µV, a4..a0 polynomial
coefficients.
6*Float
SRC 24
R/W
90000;
0; 0; 0;
0,01; 0
CUSTOM_TC_POLY_4
16
Polynomial part 4 of Custom defined TC converting µV to °C.
Consisting of: maximum input value in µV, a4..a0 polynomial
coefficients.
6*Float
SRC 24
R/W
120000;
0; 0; 0;
0,01; 0
CUSTOM_TC_POLY_5
17
Polynomial part 5 of Custom defined TC converting µV to °C.
Consisting of: maximum input value in µV, a4..a0 polynomial
coefficients.
6*Float
SRC 24
R/W
150000;
0; 0; 0;
0,01; 0
CUSTOM_TC_RJ_POLY
18
RJ Polynomial part of custom defined TC, converting °C to µV.:
a3..a0 coefficients.
4*Float
SRC 16
R/W
0;0;0;0
CUSTOM_RTD_NAME
19
Name of Custom defined RTD ( LIN_TYPE = 241)
OCTET_
STRING SRC 20
R/W
”Linear
RTD”
CUSTUM_RTD_POLY_CO
UNT
20
Number of 4. order polynomial parts for Custom defined RTD
Unsigned
8
SRC 1
R/W
5
CUSTOM_RTD_MIN_IN
21
Minimum input limit in Ohm’s for Custom defined RTD
Float
SRC 4
R/W
0
CUSTOM_RTD_MIN_OUT 22
Minimum usable output value of polynomial set for Custom
defined RTD
Float
SRC 4
R/W
0
CUSTOM_RTD_MAX_OU
T
23
Maximum useable output value of polynomial set for Custom
defined RTD
Float
SRC 4
R/W
100,00
CUSTOM_RTD_POLY_1 24
Polynomial part 1 of Custom defined RTD converting Ohm to
°C. Consisting of maximum input value in Ohms, a4..a0
polynomial coefficients.
6*Float
SRC 24
R/W
2000; 0;
0; 0;
0,01; 0
CUSTOM_RTD_POLY_2 25
Polynomial part 2 of Custom defined RTD converting Ohm to
°C. Consisting of maximum input value in Ohms, a4..a0
polynomial coefficients.
6*Float
SRC 24
R/W
4000; 0;
0; 0;
0,01; 0
CUSTOM_RTD_POLY_3 26
Polynomial part 3 of Custom defined RTD converting Ohm to
°C. Consisting of maximum input value in Ohms, a4..a0
polynomial coefficients.
6*Float
SRC 24
R/W
6000; 0;
0; 0;
0,01; 0