The Transducer Block
3.8 AI_TRANSDUCER Block (PA Slot 3) Parameter List
SITRANS TH400 PROFIBUS® PA
Configuration Manual, 02/2007, A5E01039143-01
35
3.8.9
Sensor Calibration Parameters
Parameter
Rel.
Index
PA
Description
Type
Store Size
byte
RO /
R/W Min. Max. Default
CAL_POINT_LO_1
70
The low calibration value applied to sensor 1
The value from either a calibrator or a reference equipment.
Float
SRC 4
R/W
-10
38
CAL_ACTUAL_LO_1
71
Entering any value will force the device to automatically measure
and store the actual low point value. Must be entered with the
applied CAL_POINT_LO_1 value.
Float
SRC 4
R/W
-10
38
CAL_POINT_HI_1
72
The high calibration value applied to sensor 1
The value from either a calibrator or a reference equipment.
Float
SRC 4
R/W
10
38
CAL_ ACTUAL _HI_1
73
Entering any value will force the device to automatically measure
and store the actual high point value. Must be entered with the
applied CAL_POINT_HI_1 value
Float
SRC 4
R/W
10
38
SENSOR_CAL_METHOD_
1
74
Enables or disables the last sensor calibration for sensor 1
103 = Factory trim standard (calibration values disabled)
104 = User trim standard (calibration values enabled)
Unsigned
8
SRC 1
R/W
103
SENSOR_CAL_LOC_1
75
The last location of the calibrated sensor.
OCTET_
STRING SRC 32
R/W
” ”
SENSOR_CAL_DATE_1
76
The last date on which the calibration was performed
7 *
Unsigned
8
SRC 7
R/W
0,0,0,0,
1,1,103
SENSOR_CAL_WHO_1
77
The name of the person responsible for the last sensor
calibration.
OCTET_
STRING SRC 32
R/W
” ”
CAL_POINT_LO_2
78
The low calibration value applied to sensor 2
The value from either a calibrator or a reference equipment.
Float
SRC 4
R/W
-10
38
CAL_ACTUAL_LO_2
79
Entering any value will force the device to automatically measure
and store the actual low point value. Must be entered with the
applied CAL_POINT_LO_2 value
Float
SRC 4
R/W
-10
38
CAL_POINT_HI_2
80
The high calibration value applied to sensor 2
The value from either a calibrator or a reference equipment.
Float
SRC 4
R/W
10
38
CAL_ACTUAL_HI_2
81
Entering any value will force the device to automatically measure
and store the actual high point value. Must be entered with the
applied CAL_POINT_HI_2 value
Float
SRC 4
R/W
10
38
SENSOR_CAL_
METHOD_2
82
Enables or disables the last sensor calibration for
sensor 2
103 = Factory trim standard (calibration values disabled)
104 = User trim standard (calibration values enabled)
Unsigned
8
SRC 1
R/W
103
SENSOR_CAL_LOC_2
83
The last location of the calibrated sensor
OCTET_
STRING SRC 32
R/W
» »
SENSOR_CAL_DATE_2
84
The last date on which the calibration was performed
7 *
Unsigned
8
SRC 7
R/W
0,0,0,0,
1,1,103
SENSOR_CAL_WHO_2
85
The name of the person responsible for the last sensor
calibration
OCTET_
STRING SRC 32
R/W
» »