Explosion protection
5.2 FM
SITRANS SL
Operating Instructions, 12/2010, A5E01132948-04
79
WARNING
Before opening the device wait at least two minutes after de-energizing. Make sure that
externally powered signals are also de-energized.
WARNING
Never repair the device on site!
Any components not mentioned in the spare parts list must be replaced or repaired by
certified field service technicians. Failure to do so will also result in loss of FM approval.
WARNING
Before switching on the analyzer make sure that the housing is closed and grounded.
The protective earth terminal of the SITRANS SL needs to be connected to the local
potential equalization system.
Death, personal injury and/or damage to property may result if this is not observed.
WARNING
The use of the alignment kit (A5E01000740) for aligning the sensors is not covered by the
FM certificate of the SITRANS SL analyzer.
Never use the kit in the hazardous area without permission of the plant manager (hot work
permit).
WARNING
Never open the cover of an energized SITRANS SL system in the hazardous area without
permission of the plant manager (hot work permit).
Death, personal injury and/or damage to property may result if this is not observed.
SITRANS SL FM meets all regulations specified in the present EU regulations (LVD
regulation 2006/95/EEC, EMC regulation 2004/108/EEC, ATEX regulation 94/9/EC) as well
as the regulations of the American market (FM).