Interfaces to automation systems
8.2 Modbus interface
SITRANS SL
Operating Instructions, 12/2010, A5E01132948-04
143
8.2.4.2
Enron Modbus
This section describes the Enron Modbus register group numbers for the SITRANS SL gas
analyzer.
Register numbers are used to identify specific data items to be read or written. Registers are
grouped by data type.
The following table depicts SITRANS SL Enron Modbus register group numbers.
Table 8- 7
Register groups
Register
Type
Description
3000
INTEGER
16 Bit integers
5000
LONG
32 Bit integers
7000
FLOAT
32 Bit IEEE floating point
Note
Offset
As default, for all Enron Modbus registers the offset between the register number and the
data addresses is 1. For example, register number 7001 has data address 1B59 (7001 in
decimal).
There is an optional support for an offset 0.
Modbus addresses 3000 ... 3999
Short Integer registers
Short Integer registers are read using function code 03 or set using function code 16.
Table 8- 8
Short Integer registers
Enron
Modbus
address
Access
Data Address
Description
3001
read
3000
Device Status
3002
read
3001
Component 1 – Concentration
Quality
3003
read
3002
Component 1 – Concentration
Time stamp – Day
3004
read
3003
Component 1 – Concentration
Time stamp – Hour
3005
read
3004
Component 1 – Concentration
Time stamp – Minute
3006
read
3005
Component 1 – Concentration
Time stamp – Second
3007
read
3006
Component 2 – Transmission
Quality