7ML19985QV81
SITRANS Probe LU (PROFIBUS PA) – QUICK START MANUAL
Page EN-5
m
mmmm
Englis
h
Mounting instructions
SITRANS Probe LU is available in three thread types: 2" NPT, 2" BSP, or PF2/G (BS EN ISO 228-1).
1.
Before inserting SITRANS Probe LU into its mounting connection, ensure that the threads
are of the same type to avoid damaging them.
2. Simply screw SITRANS Probe LU into the process connection and hand tighten.
Wiring
Power
1
Connecting
SITRANS Probe LU (PROFIBUS PA)
1.
Strip the cable jacket for approximately
70 mm (2.75") from the end of the PROFIBUS PA
cable, and thread the wires through the gland
2
.
2. Connect the wires to the terminal as shown:
Probe LU (PROFIBUS PA) is not polarity-sensitive.
Note:
Ideally, mount SITRANS Probe LU so that the face of the transducer is at least
300 mm (1 ft) above the highest anticipated level.
WARNINGS:
DC terminals shall be supplied from an SELV
1
source in accordance with
IEC-1010-1 Annex H.
All field wiring must have insulation suitable for rated voltages.
1.
Safety Extra Low Voltage
Note:
• For detailed wiring instructions, please see the full manual.
• For Intrinsically Safe setups (FM/CSA Class I, Div. 1), see
FM/CSA Intrinsically Safe
Connection Drawing
on page 1 of Appendix A, for drawing number 23650617.
• For Non-Incendive setups (FM: Class I, Div. 2), see
FM: Class I, Div. 2 Connection Drawing
on page 5 of Appendix A, for drawing number 23650583.
• The non-metallic enclosure does not provide a continuous ground path between conduit
connections: use grounding-type bushings and jumpers.
• Separate cables and conduits may be required to conform to standard instrumentation
wiring practices, or electrical codes.
2.
If cable is routed through conduit, use only approved suitable-size hubs for waterproof applications.
shield
wire
Содержание sitrans PROBE LU
Страница 1: ...Quick Start Manual May 2008 PROBE LU PROFIBUS PA sitrans ...
Страница 18: ...Page EN 16 SITRANS Probe LU PROFIBUS PA QUICK START MANUAL 7ML19985QV81 mmmmm English Notes ...
Страница 34: ...Side DA 16 SITRANS Probe LU PROFIBUS PA KVIKSTART MANUAL 7ML19985QV81 mmmmm Dansk Bemærkninger ...
Страница 50: ...Seite DE 16 SITRANS Probe LU PROFIBUS PA KURZANLEITUNG 7ML19985QV81 new mmmmm Deutsch Notizen ...
Страница 82: ...Página ES 16 SITRANS Probe LU PROFIBUS PA GUÍA PARA LA PUESTA EN MARCHA 7ML19985QV81 new mmmmm Español Notas ...
Страница 98: ...Page FR 16 SITRANS Probe LU PROFIBUS PA MISE EN SERVICE SIMPLIFIEE 7ML19985QV81 new mmmmm Français Notes ...
Страница 114: ...Pagina IT 16 SITRANS Probe LU PROFIBUS PA MANUALE PER L AVVIO RAPIDO 7ML19985QV81 mmmmm Italiano Annotazioni ...
Страница 130: ...Pagina NL 16 SITRANS Probe LU PROFIBUS PA QUICK START HANDLEIDING 7ML19985QV81 mmmmm Nederlands Opmerkingen ...
Страница 146: ...Página PT 16 Sonda LU SITRANS PROFIBUS PA MANUAL DE ARRANQUE RÁPIDO 7ML19985QV81 mmmmm Português Notas ...
Страница 162: ...Sivu SU 16 SITRANS Probe LU PROFIBUS PA PIKAKÄYTTÖOPAS 7ML19985QV81 mmmmm Suomi Huomautukset ...
Страница 178: ...Sida SV 16 SITRANS Probe LU PROFIBUS PA MANUAL FÖR SNABBSTART 7ML19985QV81 mmmmm Svenska Anmärkningar ...