Maintenance and servicing
5.7 Replacing parts
SITRANS P DS III
50
Service Manual, 06/2014, A5E34219855-01
4.
Use a 2.5 mm Allen key to loosen the locking screw for fixing the measuring cell.
NOTICE
Damage to the ribbon cable
To avoid damage to the ribbon cable when installing or removing the measuring cell, do
not kink, pinch, tension or twist it.
•
Do not kink, pinch, tension or twist the ribbon cable.
5.
Turn the measuring cell counter-clockwise to remove it from the housing. To do this, use
a matching open-end wrench (relative pressure transmitter) or rotate the measuring cell
by hand.
Содержание SITRANS P DS III
Страница 1: ......
Страница 2: ......
Страница 13: ...Introduction 1 8 Notes on warranty SITRANS P DS III Service Manual 06 2014 A5E34219855 01 13 ...
Страница 14: ......
Страница 18: ...Safety notes 2 3 Use in hazardous areas SITRANS P DS III 18 Service Manual 06 2014 A5E34219855 01 ...
Страница 22: ...Description 3 3 Measuring point label layout SITRANS P DS III 22 Service Manual 06 2014 A5E34219855 01 ...
Страница 56: ...Technical specifications SITRANS P DS III 56 Service Manual 06 2014 A5E34219855 01 ...
Страница 57: ...SITRANS P DS III Service Manual 06 2014 A5E34219855 01 57 Ordering data for spare parts accessories 7 ...
Страница 58: ...Ordering data for spare parts accessories SITRANS P DS III 58 Service Manual 06 2014 A5E34219855 01 ...
Страница 59: ...Ordering data for spare parts accessories SITRANS P DS III Service Manual 06 2014 A5E34219855 01 59 ...
Страница 60: ...Ordering data for spare parts accessories SITRANS P DS III 60 Service Manual 06 2014 A5E34219855 01 ...
Страница 61: ...Ordering data for spare parts accessories SITRANS P DS III Service Manual 06 2014 A5E34219855 01 61 ...
Страница 63: ...Ordering data for spare parts accessories SITRANS P DS III Service Manual 06 2014 A5E34219855 01 63 ...
Страница 64: ......
Страница 74: ......