Maintenance and servicing
5.6 User data following replacement
SITRANS P DS III
Service Manual, 06/2014, A5E34219855-01
35
5.6
User data following replacement
This chapter describes which user data is reset to the factory setting following replacement
of the application electronics or measuring cell.
The following tables are identified as follows:
*: The parameter names identified by * can be found in SIMATIC PDM under the menu items
"Device" or "View".
①
: The user data identified by
①
is reset to the factory setting following replacement of the
measuring cell.
②
: The user data identified by
②
is reset to the factory setting following replacement of the
application electronics.
Table 5- 1
SITRANS P DS III with HART
Parameter name in SIMATIC PDM
Status of user data following
replacement of application
electronics or measuring cell
Value following resetting to factory
setting
Operating unit / TAG
①
-
Device / sensor serial number
①
-
Unit and cycle time / unit
①
bar
Unit and cycle time / cycle time
①
90 ms
Measuring limits / min. measuring range
①
-
Measuring limits / min. start-of-scale value,
max. full-scale value
①
-
Measured-value scaling, start-of-scale value
①
0.0
Measured-value scaling, full-scale value
①
Upper sensor limit
Analog output limits / lower/upper limit
①
3.8 / 20.5 mA
Alarm status / analog output alarm type
①
Lower
Alarm status / lower/upper alarm value
①
3.6 / 22.8 mA
Damping
①
0.1 s
Characteristic / transfer function
①
Linear
Characteristic / start of root extraction
①
5 %
Display and user interface / tracking unit
①
mA
Design (all material data)
①
-
Sensor calibration / upper/lower calibration
point
①
Like factory-set sensor limits, without
transformation
Sensor calibration / lower calibration point
①
0.0
Sensor calibration / upper calibration point
①
Upper sensor limit
Service interval for sensor (all settings)
①
-
Содержание SITRANS P DS III
Страница 1: ......
Страница 2: ......
Страница 13: ...Introduction 1 8 Notes on warranty SITRANS P DS III Service Manual 06 2014 A5E34219855 01 13 ...
Страница 14: ......
Страница 18: ...Safety notes 2 3 Use in hazardous areas SITRANS P DS III 18 Service Manual 06 2014 A5E34219855 01 ...
Страница 22: ...Description 3 3 Measuring point label layout SITRANS P DS III 22 Service Manual 06 2014 A5E34219855 01 ...
Страница 56: ...Technical specifications SITRANS P DS III 56 Service Manual 06 2014 A5E34219855 01 ...
Страница 57: ...SITRANS P DS III Service Manual 06 2014 A5E34219855 01 57 Ordering data for spare parts accessories 7 ...
Страница 58: ...Ordering data for spare parts accessories SITRANS P DS III 58 Service Manual 06 2014 A5E34219855 01 ...
Страница 59: ...Ordering data for spare parts accessories SITRANS P DS III Service Manual 06 2014 A5E34219855 01 59 ...
Страница 60: ...Ordering data for spare parts accessories SITRANS P DS III 60 Service Manual 06 2014 A5E34219855 01 ...
Страница 61: ...Ordering data for spare parts accessories SITRANS P DS III Service Manual 06 2014 A5E34219855 01 61 ...
Страница 63: ...Ordering data for spare parts accessories SITRANS P DS III Service Manual 06 2014 A5E34219855 01 63 ...
Страница 64: ......
Страница 74: ......