2
SFIDK.PS.029.Q3.02
SITRANS F US SONOFLO
®
®
®
®
®
SONO 3000
SONO 3000 Ex-d
Signal converters
Enclosure
IP 67
19" insert
19" insert
IP 67
IP 00
IP 65
Mounting
Compact or separate
19" rack wall mounting and
Compact or separate
wall mounting
panel mounting
wall mounting
Accuracy (error limit)
0.5% of actual flow
Materials
Polyamid
ALU
ABS Plastic
ASTM 316
Outputs
1 current
3 current
3 current
1 current
1 frq./pulse
2 frq./pulse
2 frq./pulse
1 frq./pulse
1 relay
2 relays
2 relays
1 relay
Display/counter
Alphanumeric: flow, quantity, error, mass flow, sound velocity, etc.
Language
English, German, French
Approval
Can be used with any EEx-approved sensor SONO 3100
and SONOKIT, when the SONO 3000
EEx de
[
ia/ib
]
IIC T6
is placed in safe area
Supply voltage
115/230 V a.c.
24 V a.c./d.c.
24 V a.c./d.c.
Sensor
Size
[[[[[
mm
]]]]]
DN 50 - DN 300
DN 100 - DN 1200
DN 200 - DN 4000
Connection examples
DIN 2501
ANSI B 16.5 150 lbs RF
None
ANSI B 16.5 300 lbs RF
Pressure
[[[[[
bar
]]]]]
PN 10, PN 16,
PN 6, PN 10, PN 16,
PN 40
PN 25, PN 40
Max. PN 40
Class 150, Class 300
Class 150, Class 300
Media temperature
max.
[[[[[
°C
]]]]]
160
200
200
Transducer
SONO 3200
SONO 3200
Integrated transducers
O-ring type
O-ring type
Flange transducers
Enclosure
IP 67
IP 68
IP 68
Approvals
EEx d II C T3-T6
EEx d IIC T2-T6
EEx d IIC T3-T6
SONO 3100
SONO 3300
SONOKIT
Stainless and
mild steel
Extended program
Mild steel
Standard program
Siemens Flow Instru-
ments range of ultra-
sonic flowmeters