Alarm, error, and system messages
8.2 List of error numbers
MAG 6000 I / MAG 6000 I Ex de transmitter
Operating Instructions, 04/2016, A5E02083319-AG
53
8.2
List of error numbers
Error
No.
Error text
Remedy text
Comment
Output
status
Input
status
1
I1 - Power on
OK
Device powered on
Active
Active
2
I2 - Add-on module
Applied
A new module has been applied to the system
Active
Active
3
I3 - Add-on module
Install
An add-on module is defect or has been removed.
This can be an internal add-on module
Active
Active
4
I4 - Param. corrected
OK
A less vital parameter in the transmitter has been
replaced by its default value
Active
Active
20
W20 - Totalizer 1
Reset manually
During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the
saved totalizer value anymore. The totalizer value
must be reset manually in order to rely on future
readings
Active
Active
20
W20 - Totalizer 2
Reset manually
During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the
saved totalizer value anymore. The totalizer value
must be reset manually in order to rely on future
readings
Active
Active
21
W21 - Pulse overflow
Adj. pulse settings
Actual flow is too big compared with pulse width and
volume/pulse
Reduced pulse
width
Active
22
W22 - Batch timeout
Check installation
Duration of batching has exceeded a predefined
maximum time
Batch output
on zero
Active
23
W23 - Batch overrun
Check installation
Batch volume has exceeded a predefined maximum
overrun volume
Batch output
on zero
Active
24
W24 - Batch neg. flow
Check flow direction
Negative flow direction during batch
Active
Active
30
W30 - Overflow
Adj. Q
max
Flow is above Q
max
settings
Max. 120 %
Active
31
W31 - Empty pipe
Pipe is empty
Zero
Active
40
P40 - SENSORPROM®
Insert/change
SENSORPROM
®
unit not installed
Active
Active
41
P41 - Parameter range
Switch off and on
A parameter is out of range. The parameter could
not be replaced by its default value. The error will
disappear at the next power-on
Active
Active