Technical Data
8
Pollution degree
2
Overvoltage category
II
Fuses information
315 mAT (for mains)
2.0 AT (for others)
Supply voltage
1)
● 11 to 30 V DC
● 11 to 24 V AC, 50/60 Hz
● 115/230 V AC, 50/60 Hz, +10% −15%
Power consumption
< 30VA, incl. MAG 5000/6000 transmitter
Output
RS 232 galvanically isolated
Enclosure rating
IP20 open briefcase / IP42 closed briefcase
Temperature
● Ambient: −5 to 45 °C
● Media: 0 to 60 °C
● During storage: −40 to 70 °C
Verification accuracy
● Sensor: Magnetic field properties < ±2%
● Insulation; No degradation on system accuracy
● Transmitter: No degradation on system accuracy
PC software
PC with at least 486 processor and Windows 95, 98, NT or XP
CE-confirmity
● Emission: EN 50081-1
● Immunity: EN 50082-2
2)
Traceability
Calibration at Siemens Hagenau (COFRAC acredited laborato‐
ries) traceable to national and international standards
Recalibration
Mandatory recalibration with one year interval to maintain tracea‐
bility
Tests/approvals
Water Research Centre, England
Limitation
Cannot be used on sensors in Ex areas, max. liquid temperature
60 °C
1)
The Verificator automatically detects and sets itself to the correct supply voltage. A 115/230 V
transmitter cannot be verified with 11 to 24 V AC/DC supply.
2)
EN 50082-1 when power supply is established through power cables connected to front of Verificator.
WARNING
Connection
Mains Power cables connected to the Verificator may not exceed 2.5 meter length and shall
be properly grounded.
SITRANS F M Verificator
Operating Instructions, 07/2019, A5E45134218-AB
37