Connecting
SITRANS F C MC2
24
Operating Instructions, 10/2009, A5E02154544-02
Connection in hazardous locations
The MC2 sensors are ex certified products (increased safety) up to 60 V.
This means that:
● All secondary circuits/terminals can be considered as limited circuits with non-hazardous
voltage of max. 60 V.
● The MASS 6000 and MC2 can be connected with no further approvals.
● Equipment connected to the signal outputs (current, pulse etc) must however obtain the
same high degree of isolation from mains supply as the MASS 6000 (Class 2 transformer
or Class 1)
See also
For further information on the electrical connection of the sensor to the transmitter, see the
operating instructions for the respective transmitters.