Connecting
SITRANS F C MC2
22
Operating Instructions, 10/2009, A5E02154544-02
Grounding terminal
Connect grounding terminal (PE) to protective earth
Non-Ex
Ex
NOTICE
In compliance to the EMC-standard / low voltage guideline, the terminal box, additional to
the meter tube of the flow sensor, has to be grounded. To make this connection, make use
of the included green/yellow cable:
•
Cut the enclosed cable lug to the shortest required length
•
Crimp the matching cable lug onto the cable
•
Join both ends and make sure both screws are securely fastened
Explosion proof instruments should fundamentally be integrated into equipotential bonding
Electrical connection
WARNING
Use in hazardous locations
Special requirements apply for the power supply, in- and output signals and ground
connections.