Modbus RTU RS-485
28
SFIDK.PS.023.D2.02
APPENDIX C
Modbus communication
examples
01 (0x01) Read coil status
Query
Slave address
1 byte
Function
1 byte
Starting Address Hi
1 byte
Starting Address Lo
1 byte
No. of Points Hi
1 byte
No. of Points Lo
1 byte
CRC
2 bytes
Slave address
1 byte
Function
1 byte
Byte Count
1 byte
Data Coil
1 byte
:
:
Data Coil
1 byte
CRC
2 bytes
Response
03 (0x03) Read holding registers
Query
Slave address
1 byte
Function
1 byte
Starting Address Hi
1 byte
Starting Address Lo
1 byte
No. of Points Hi
1 byte
No. of Points Lo
1 byte
CRC
2 bytes
Slave address
1 byte
Function
1 byte
Byte Count
1 byte
Data Hi
1 byte
Data Lo
1 byte
:
:
Data Hi
1 byte
Data Lo
1 byte
CRC
2 bytes
Response
Examples
Read absolute mass flow (4:03001)
Query:
1,3,11,184,0,2,70,10
Response:
1,3,4,64,195,82,139,98,200
Absolute mass flow =
6.10383 kg/sec.
Read absolute density (4:03005)
Query:
1,3,11,188,0,2,7,203
Response:
1,3,4,68,54,162,145,182,1
Absolute density =
730.54010 kg/m
3
Read sensor temperature (4:03007)
Query:
1,3,11,190,0,2,166,11
Response:
1,3,4,189,216,115,34,251,77
Sensor temperature =
−
0.10569
°
C
Example:
Read Hold Totalizer 1 (0:00011)
Query:
1,1,0,10,0,1,221,200
Response:
1,1,1,0,81,136
APPENDIX C