Modbus RTU RS-485
25
SFIDK.PS.023.D2.02
Modbus
Modbus
No. of
Data
Read/
register
address
bytes
type
Description
write
4:03201
3200
2
Word
System status:
bit0=RunIndicator
bit1=ErrorPendingOn
bit2=ErrPendChanged
bit3=Batching
R
bit4=BatchPaused
bit5=Zeroadjusting
bit6=Totalizer1Hold
bit7=Totalizer2Hold
4:03202
3201
2
Word
Error pending 1
*
(oldest/first error)
0xFFFF = no error
R
4:03203
3202
2
Word
Error pending 2
*
R
4:03204
3203
2
Word
Error pending 3
*
R
4:03205
3204
2
Word
Error pending 4
*
R
4:03206
3205
2
Word
Error pending 5
*
R
4:03207
3206
2
Word
Error pending 6
*
R
4:03208
3207
2
Word
Error pending 7
*
R
4:03209
3208
2
Word
Error pending 8
*
R
4:03210
3209
2
Word
Error pending 9
*
R
4:03211
3210
2
Byte
Batch status
0 = Idle
1 = Ready
R
2 = Running
3 = Paused
Modbus
Modbus
No. of
Data
Read/
register
address
bytes
type
Description
write
4:03212
3211
2
Byte
Zero adjust progress
0 = Idle
1 = Ready
R
2 = Running
Zero adjust progress MAG
6.
Modbus holding registers
Modbus
Modbus
No. of
Data
Read/
register
address
bytes
type
Description
write
4:03212
3211
4
Float
Zero adjust progress (0-100 %)
R
Zero adjust progress MASS
6.12 Flowmeter status
*
Further information about the
Error pending numbers
can be found in the MAG/MASS 6000 manuals.