Area dose product measurement device
5
Siemens AG
SPR2-230.033.04
Page 5 of 16
SIREMOBIL Iso-C
Medical Engineering
Rev. 03
06.05
CS PS 24
Safety Information, Risk of Infection
1
Danger of injuries on mechanical parts!
Non-compliance can lead to slight to medium injuries, especially
of the hands.
Before starting service or maintenance work, check the system by
a visual inspection for mechanical safety defects which in the
case of inattention could lead to injuries. For example, projecting,
bent or broken parts / covers, damaged plugs or cables.
Remove existing safety defects by replacement or repair of the
damaged parts.
Perform the particular work steps with special care and attention
to detail.
If needed, wear working gloves.
Danger of burns on hot parts or components!
Non-compliance can lead to slight to medium injuries, especially
of the hands.
After opening the cover panels, parts and components (e.g.
power components, cooling units, electromagnetic brakes) are
accessible that can have temperature of > 50C during operation.
To avoid burns, switch the system off prior to touching parts or
components and let them cool at least 5 minutes.
Danger of infections due to pathogens!
Non-compliance can lead to severe injuries and even death.
This product can be contaminated by infected blood or other
bodily fluids.
Avoid all contact with blood or other bodily fluids!
Strictly observe the safety information in ARTD-002.731.37...
regarding prevention of infectious diseases during customer ser-
vice calls.
CAUTION
CAUTION
WARNING