SIREMOBIL Iso-C
SPR2-230.033.04
Page 10 of 16
Siemens AG
Rev. 03
06.05
CS PS 24
Medical Engineering
10
Area dose product measurement device
Memoskop Programming
•
Turn on the SIREMOBIL and wait for the start-up.
•
Insert the Memoskop parameter disk (see System folder Register 10) in drive A of the
service PC.
•
Start the Memoskop service program.
•
Click the "Get from Memoskop" button in the main menu. The parameters are transferred
from the Memoskop to the Service PC.
•
Select the "Others" menu.
•
Select "Disable dose rate selection in User Set Up". The Check button is not marked.
•
Select "Storage of dose on disk". The Check button is marked.
•
Click on the "Back" button. The "Others" window closes.
•
Select the "Hardcopy Setup" menu.
•
Select "Dose on Hardcopy output". The Check button is marked.
•
Click on the "Back" button. The "Hardcopy Setup" window closes.
•
Click on the "Put to Memoskop" button in the main menu. The parameters are transfer-
red from the service PC to the Memoskop and saved.
•
Select the "File" menu.
•
Select the sub-menu "Save" and save the parameters on disk.
•
Exit the Memoskop service program.
•
Turn the SIREMOBIL off and the on again to reset the serial interface.
After selecting the "Disable dose rate selection in User Set Up"
parameter, the "Storage of dose on disk" and "Dose on hardcopy
output" parameters become available in the Memoskop User
Setup and visible to the customer.
NOTE