Operation and configuration
8.3 Additional drivers and applications
SIMATIC HMI IPC677C
82
Operating Instructions, 04/2013, A5E02722710-04
8.3.1
Standard calibration procedure
Procedure
1.
Select "Start > Programs > UPDD > Settings".
The dialog box "UPDD Console" opens.
Figure 8-1
Standard calibration
2.
Select the Controller
①
you wish to calibrate.
3.
Click on the "Calibration" tab
②
.
4.
Activate the option box "Number of points" with the 25 points calibration
③
.
5.
Click the button "Calibrate"
④
.
The calibration screen is displayed in the selected display.
6.
Touch the corresponding selections one after the other.
The entry is confirmed by a check mark, the next selection is displayed.
7.
Confirm all input prompts (arrows, or crosses in the center) until the complete screen has
been calibrated.
8.
Finally, confirm the input prompt "Confirm".