![Siemens Milltronics BW100 Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/siemens/milltronics-bw100/milltronics-bw100_instruction-manual_374197038.webp)
Page 34
Milltronics BW100 - INSTRUCTION MANUAL
7ML19985DJ01
mmm
mm
Re
ca
libr
ation
Direct Zero
Use direct zero entry (P367) when replacing software or hardware, if it is not convenient
to perform an initial zero. A record of the last valid zero count is required.
Refer to
Maneuvering
on page 18 for parameter selection and value change.
%
current zero count
enter previously recorded
zero count
new zero count accepted