Page 22
Milltronics BW100 - INSTRUCTION MANUAL
7ML19985DJ01
mmm
mm
St
art
Up
Place the BW100 into the PROGRAM mode at P291.
Remove the test weight, insure that the BW100 SW2 is in ‘close’ position and that P1 is
left as set.
Note:
Performing a balance procedure requires a subsequent zero and span
recalibration.
test weight
test
weight
test weight on
load cell B side
select P291 ensure SW2
is closed
enter and auto advance to P292,
open SW2
enter and auto advance to P293,
leave SW2 open
test weight on
load cell A side
SW2
SW2
SW2
SW2
P1
enter and auto advance to P294,
close SW2
enter and auto advance to P295
adjust until
## = 0 +/- 5
closed
closed
open
open
’B’
’B’
’A’
’A’