Milltronics BW100 - INSTRUCTION MANUAL
Page 31
mmm
mm
Recalibr
ati
on
If the span adjust value is within the accuracy requirements of the weighing system, the
material test was successful and normal operation can be resumed.
If the span adjust value is not acceptable, repeat the material test to verify repeatability. If
the result of the second material test differs considerably, consult Milltronics or its agent.
If the span adjust values are significant and repeatable, perform a manual span adjust:
Design Changes
Where parameters have been changed with a resultant impact on the calibration or do
not take effect until a recalibration is done, the parameter warning icon is displayed. In
order to clear the icon, perform a zero and span recalibration after the reprogramming
session is complete.
If significant changes have been made, an initial zero (P377) and/or initial span (P388)
may be required.
Recalibration
In order to maintain the accuracy of the weighing system, periodic zero and span
recalibrations are required. Recalibration requirements are highly dependent upon the
severity of the application. Perform frequent checks initially, then as time and experience
dictate, the frequency of these checks may be reduced. Record deviations for reference.
PROGRAM
mode,
P019
enter %
span adjust,
e.g. 0.05
jumps to P017 adjusted
test load, e.g. 97.95
refer to
Maneuvering
,
page 18 for
parameter selection
and value change
if % span adjust is
negative, be sure
to enter as
negative value
Verify the results of the span adjust by material test or return to normal operation