2 System
MAMMOMAT Balance
SPB7-115.831.01 Page 9 of 22
Siemens AG
Maintenance Instructions
Rev.01 05.04
Medical Solutions
2 System
SIM
Radiation protection
•
Check the radiation protection shield for possible damage (n°1, see figure).
•
Information regarding the lead equivalent value must be legible.
SIM
C-Arm – attachment and handles
•
Check by pulling/pushing on both handles that no parts have got loosened (n°2, see figure).
SIM
Tube cover
•
Check that tube cover hasn’t got loosened by pulling/pushing it in different directions (n°3, see
figure).
SIM
Table retain
•
Check that all tables are retained when inserted and can’t be removed with very light pulling force.
•
Inspect connectors and connector cover integrity on the table support (n°4, see figure).
SIE
Cable
•
System OFF
•
Check the condition of the power connection cable and the corrugated tubing (n°5, see figure).
•
Check the condition of foot pedal switches (n°6, see figure).
PMF Rotation movements
•
Is the rotation smooth and noiseless through the entire range?
•
Is the rotation brake holding the C-Arm position?
•
Is the mechanical goniometer functional?
•
Is the C-Arm rotation disabled without table?