5 Rotating anode starter
MAMMOMAT Balance
SPB7-115.831.01 Page 12 of 22
Siemens AG
Maintenance Instructions
Rev.01 05.04
Medical Solutions
5 Rotating anode starter
PMF Protection circuit check
•
For protection circuit check, the anode must be put in rotation and it is advisable as follows:
DEMO mode with DIP 3 SW1 of board 01-170 switched ON
Manual mode with at least 200 mAs
In this mode, no X-rays are generated and measurement time is sufficiently long.
PMF Control LED (brake enabled)
•
PCB 92-074
LED1
(yellow) bright during normal rotation phase (RUN)
LED2
(yellow) bright during breaking or alarm
LED3
(yellow) bright during breaking
LED5
(green) bright in stand by or rotation
LED6
(green) bright during rotation
V(TP3) = 0.65xV(TP1)
V(TP4) = 0.65xV (TP2)
Stand by
Tube run
Stand by