7 High tension power supply
MAMMOMAT Balance
SPB7-115.831.01 Page 15 of 22
Siemens AG
Maintenance Instructions
Rev.01 05.04
Medical Solutions
7 High tension power supply
PMF kV waveforms
The oscillogram shows the H.V.vaweform at 20 kV and 35 kV measured on TP1.
kV wave form has no overshoot for any kV / mA loading factors combination.
kV risetime
PMF Internal kV check (TP1/TP2)
•
DVM 1000Vdc f.s. or oscilloscope with 1:100 1000Vdc probe can be connected to TP1/TP2
(GND) of PCB 94-132, see page 17.
•
Remove any table, select Manual Mode and an mAs value high enough to allow stable