Preparatory work
3 - 3
Siemens-Elema AB
Register 03
SP B7-230.033.07
Page 3 of 8
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Removing the transport safeguards
3
Removing the swivel-arm system transport safeguard
3
1.
Loosen the transport safeguard nuts (1/Fig. 2) and remove the tube head screws
(2/Fig. 2).
2.
Push the red transport safeguard (3/Fig. 2) inwards, away from the tube head.
Transport safeguard for the lifting carriage
3
1.
Loosen the two screws (1/Fig. 3) and remove the red transport safeguard (2/Fig.
3). Reinsert screws (1/Fig. 3), they might be needed for service purposes.
Fig. 2
Swivel-arm system transport safeguard
Removal of the swivel-arm system transport safeguard is done
after the equipment has been powered up, see further section
"Attaching the swivel-arm covers" on page 7 - 1.
Fig. 3
Transport safeguard for the lifting carriage
MAM00081
1
2
3
NOTICE
2
1
M
A
M
0008
2