7 - 1
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 1 of 2
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Attaching the swivel-arm covers
7
Attaching the swivel-arm covers
7
Arranging the swivel-arm system
7
1.
Remove the cable tie securing the balancing spring at the back of the stand.
2.
Install object table, no compression force (swivel-arm system still upside down).
3.
Remove the protective strips from the metal curtain.
4.
Run the lifting carriage upwards.
5.
Put back the protective strips onto the metal curtain.
6.
Rotate the swivel-arm system to 0 degrees by pressing one of the switches for
clockwise rotation on the supporting arm.
7.
System and mains "OFF".
8.
Remove the swivel-arm system transport safeguard, see Chapter 3 "Preparatory
work".
Connecting the cables to control-button boards and patient handles
7
Before installing the side covers, the cables are to be connected.
MAMMOMAT 1000/3000
1.
Connect cables X807 (1) to control-button circuit boards D807 (2).
2.
Connect ground wires (3) to patient handles (4).
Fig. 1
Connecting the cables
MAM00108
2
1
4
3
2
1
4
3
MAMMOMAT 1000/3000